3. When the codes match, press the
Mode button. A blinking blood drop
shows that the monitor is ready for you
to test. To run a control solution test,
refer to page 16. To run a blood glucose
test, refer to page 21.
To turn off the monitor, remove and
discard the used test strip. The monitor
turns off automatically.
Getting Started • 15
14 •
Getting Started
Running a Control
Solution Test
Finish Coding
WHY RUNNING A CONTROL SOLUTION TEST
IS IMPORTANT
The control solution test confirms that your monitor and test strips are
working correctly. A control solution test is similar to a blood glucose
test, except you use BD™ Control Solution and not a blood sample
.
You Should Run a Control Solution Test:
Before using your monitor for the first time and at least once
a week thereafter.
•Each time you open a new box of BD™ Test Strips.
If you leave the test strip vial cap open.
If you drop your monitor.
If your results do not match the way you feel, or you think
your results are higher or lower than expected.
•To check the performance of the monitor and test strips.
CAUTION:
The BD™ Control Solution range printed on the
test strip vial is for control solution only. It is used to test the
performance of the monitor and test strip. It is not a recommended
range for your blood glucose level.
Important Control
Solution Information
•Use only the BD™ Control Solution for the test.
Check the expiration date on the control solution vial.
Do not use the control solution if the expiration date has
passed.
Store only for 3 months after first opening. When you open a
new vial of control solution, count forward 3 months and write
that date on the label of the control solution vial. Discard any
remaining solution after the date you have written on the vial.
Store the control solution tightly closed at room temperature
below 30°C (86°F). Do not refrigerate or freeze.
Shake control solution well before using.